DISPLAY
:
4-4/5 digits 50,000 counts. Selectable stable mode 5-4/5 digits 500,000 counts for DC voltage, 6 digits 999,999 counts for Hz
POLARITY
:
Automatic
UPDATE RATE
:
4-4/5 digits fast mode : 5 per second nominal;5-4/5 digits stable mode : 1.25 per second nominal; 42 Segments Bar graph : 60 per second max.
LOW BATTERY
:
Below approx. 7V
OPERATING TEMPERATURE
:
0oC to 45oC
RELATIVE HUMIDITY
:
Maximum relative humidity 80% R.H. For temperature upto 31oC decreasing linearly to 50% relative humidity at 45oC
POLLUTION DEGREE
:
2
STORAGE TEMPERATURE
:
-20oC to 60oC, < 80% R.H. (With battery removed)
ALTITUDE
:
Operating below 2000m
TEMPERATURE COEFFICIENT
:
Nominal 0.1 x (specified accuracy) / oC @
(0oC -- 18oC or 28oC -- 40oC), or otherwise specified.
SENSING
:
AC, AC + DC True RMS
POWER SUPPLY
:
Single Alkaline 9V battery.
POWER CONSUMPTION
:
5mA typical
APO TIMING
:
Idle for 4.5 minutes
APO CONSUMPTION
:
20µA
DIMENSION
:
186(L) mm x 87(W) mm x 35.5(H) mm;
198(L) mm x 97(W) mm x 55(H) mm with Holster
WEIGHT
:
390 gm, 500 gm with Holster
SPECIAL FEATURES
:
Record Max, MIN, MAX-MIN readings. Crest (Instantaneous Pead Hold) MAX, MIN, MAX-MIN readings. Relative zero mode. 500,000 counts high resolution stable reading mode. Backlighted display dBm readings. T1-T2 differential temperature readings %4-20 mA loop current readings. High noise rejection filtered Line Level Frequency mode. Data Hold. Audible & Visible input warning.
STANDARD ACCESSORIES
:
Test leads (pair), holster, Battery installed, User’s manual,
K-type thermocouple x 1.
OPTIONAL ACCESSORIES
:
PC interface Kit (RS232 optical adapter cable + RS232 Software CD) , pins to type-K socket plug adapter.


 

The meter is protected, against the users, by double insulation per EN61010-1 and IEC61010-1 2nd Edition (2001) to CAT III 1000V & CAT IV 600V. The meter also meets UL3111-1 (1994) * and CSA C22.2 No. 1010-1-92* to CAT III 1000V *Category IV safety standard was first released in IEC61010-1 2nd Edition in year 2001, and was yet an UL published standard at the time this catalog was written.
E.M.C.
:
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995) and
EN61000-4-3(1996)
OVERLOAD PROTECTIONS
:
µA & mA : 1A/1kV(or 0.44A/1kV)IR 10kA, Ffuse
A :15A/1kV, IR 10kA(or 11A/1kV,IR 20kA), F fuse
 V : 1050 Vrms, 1450 V peak
mV, Ω & Others : 1050 Vrms, 1450 V peak
TRANSIENT PROTECTION
:
8kV lightning surge (1.2/50µs)